234 BD FACSCanto II Instructions for Use
Cytometer Specifications
Dimensions Height: 64 cm (25.2 in.)
Width: 91 cm (35.7 in.)
Depth: 61 cm (24 in.)
Workspace dimensions Height (with flow cell access door open): 85 cm
(33.5 in.)
Unit designed to fit lab bench 55.9 cm (22 in.)
depth.
Operational clearances, cytometer Left side: 30 cm (11.8 in.) between unit and other
objects or wall to permit proper air flow and access
to the main power button and circuit breaker
Right side: 30 cm (11.8 in.) between unit and other
objects or wall to permit proper air flow
Top: 22.5 cm (8.9 in.) between unit and other
objects or wall to permit opening of flow cell access
door
Weight 146 kg (320 lb)—cytometer only, excluding
Loader and computer
Maximum 168 kg (370 lb)—including Loader
Power requirements 100/115/230 VAC (50–60 Hz)
Current:
5A at 115 VAC
2.5A at 230 VAC
Power consumption 500 W