188 BD FACSCanto II Instructions for Use
6 Replace both fuses.
Ensure that the fuses are positioned as shown.
7 Make sure the text for your countrys voltage is right-side up.
8 Slide the drawer back into the cytometer until it snaps into place.
9 Close the fuse access door.
10 Reconnect the power cord to the fluidics cart.
11 Plug the cytometer power cord into the wall outlet and switch on the power.
For protection against risk of fire, replace fuses only with those provided by
BD Biosciences.
right wrong