204 BD FACSCanto II Instructions for Use
A communication error
was encountered
Communication problem
between hardware and
software
1 Turn on the power.
2 Connect the Ethernet cable to
the cytometer and computer.
3 Shut down the software,
computer, and cytometer, and
restart them.
Fluidics cart off or
disconnected
Switch on the fluidics cart circuit
breaker; make sure the power
cable is connected at both ends.
Cytometer setup was
aborted by user
Setup aborted Perform setup again.
Cytometer setup was
aborted because the
loader doors were
opened
Loader doors opened Perform setup again, keeping
loader doors closed.
There is a vacuum error Vacuum tubing to waste
cart disconnected
Reconnect waste tank or tubing,
and remove kinks.
Waste tank disconnected,
waste tubing
disconnected or pinched
Clogged aspirator arm Call BD Biosciences.