Chapter 5: Cytometer QC and Setup 59
7 Unload the beads tube when prompted, using these steps:
Hold the sample tube while pushing the aspirator arm to the left.
Remove the tube from the SIT.
Release the aspirator arm.
SIT cleaning occurs when the aspirator arm comes to center.
8 Decide what to do next:
If setup is unsuccessful or if setup results are out of range, do not click
Finish. Note the message provided by the software (eg, Figure 5-5 on
page 60) and consult BD FACSCanto Clinical Software
Troubleshooting on page 199.
If setup is successful, see Table 5-1 on page 60.
NOTICE It is important to monitor setup data for trends. See Reviewing
Levey-Jennings Reports on page 74.
Incorrect values are saved if you accept a setup that is unsuccessful or setup
results that are out of range.