68 BD FACSCanto II Instructions for Use
7 If there are more tubes, click , and then .
If this is the last tube, proceed to step 11 on page 69.
(Manual only) Remove the current tube when prompted, using the
following method:
• When the aspirator arm lowers, hold your sample tube while you push
the aspirator arm to the left.
• Remove the tube from the SIT.
• Release the aspirator arm.
SIT cleaning occurs when the aspirator arm comes to center.
NOTICE To prevent backflow into the tube, follow the tube removal
sequence exactly. For more information, see Cytometer Troubleshooting on
page 190.
8 (Manual only) When prompted, place the next tube on the SIT, and click
.
9 Repeat steps 3 through 6.
If you move the arm without holding the tube, the tube could fall off the
SIT and expose you to potentially biohazardous sample.