EasyManua.ls Logo

BeneVision N1 - Environmental Requirements

Default Icon
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-2
Working
environment
0~40 15%~95% 57.3~105.3
Storage and
transport
environment
–20~60 10%~95% 57.3~105.3
Sidestream CO
2
module
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Working
environment
5~40 15%~95% 57.3~105.3
Storage and
transport
environment
–20~60 10%~95% 57.0~107.4
Mainstream CO
2
module
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Working
environment
0~40 10%~90% 57.0~107.4
Storage and
transport
environment
–20~60 10%~90% 53.3~107.4
PiCCO module
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Working
environment
10~40 15%~75% 57.0~107.4
Storage and
transport
environment
–20~60 10%~90% 16.0~107.4
N1 Dock and N1 Modular Rack
Item Temperature (°C) Relative Humidity (Non-condensing) Atmospheric Pressure (kPa)
Working
environment
0~40 15%~95% 57.0~107.4
Storage and
transport
environment
–20~60 10%~95% 16.0~107.4
NOTE
Unless otherwise specified, the environmental specifications of a parameter module are the same as
those of the main unit.

Table of Contents

Related product manuals