5-16
monitor to one in the same network segment as that of the network printer (refer to the instructions for use of the
printer).
2) Use current N1 net setting: Select Main Menu→Maintenance→User Maintenance→ enter the password → Dock
Setup. Set Net Setting Type to Use current N1 net setting. Select Main Menu → Maintenance >>→ User
Maintenance >>→ enter the password → Network Setup. Set the IP address of the monitor to one in the same
network segment as that of the network printer (refer to the instructions for use of the printer).
Combination 2: N1 alone
1) Select Main Menu→Maintenance >>→User Maintenance >>→ enter the password →Network Setup. Set the IP
address of the monitor to one in the same network segment as that of the network printer (refer to the instructions
for use of the printer).
3 Set the printer:
Combination 1: N1+Dock
1) Host mode: Select Main Menu→Maintenance→User Maintenance→ enter the password →Dock Setup. Set
Work Mode to Host Mode. Select the printer setup tab. Set the IP address of the printer to one in the same
network segment as that of the monitor. Select Print Test Page to verify that the printer works properly.
2) Dock mode: Select Main Menu→Maintenance→User Maintenance→ enter the password →Dock Setup. Set
Work Mode to Dock Mode. On the User Maintenance interface, select Print→Printer, and set the IP address of
the printer. Select Print Test Page to verify that the printer works properly.
Combination 2: N1 alone
1) Select Main Menu→Maintenance→User Maintenance→enter the password →Print→Printer. Set the IP address
of the printer. Select Print Test Page to verify that the printer works properly.
5.7.2 Print Function Test
1 Enter the Demo mode of the monitor.
2 Select Main Menu→Report column →Normal Report→Realtime Report. Click Print. The network printer prints
the real-time report.
5.8 Battery Check
5.8.1.1 Functionality Test
1. Remove the battery from the monitor.
2. Check that the monitor works properly when being powered by the external DC power supply or by the AC power
supply through the Dock.
3. Install the battery according to the procedures described in the Operator's Manual.
4. Disconnect the external DC power supply or the AC power supply on the Dock. Check whether the monitor still
works properly.
5.8.1.2 Performance Test
Refer to the related content in chapter 21 Battery in BeneView N1 Patient Monitor Operator's Manual.