8-10
4. Loosen the NIBP pump connection cable and NIBP valve connection cable.
5. Loosen the NIBP air tubing on the parameter board, and remove the parameter board.
8.3.7 Removing the NIBP Gas Circuit and NIBP Valve
1. Loosen the connection between the NIBP gas circuit and interface board.
2. Loosen the connection between the NIBP gas circuit and integrated pump pipe.
3. Loosen the two M3×5 screws fixing the valve frame.
4. Remove the NIBP valve.