EasyManua.ls Logo

BeneVision N1 - 5.2.3.2 Accuracy Test; 5.2.3.3 Module Calibration

Default Icon
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-5
Calibrate to check that the current CO
2
flow rate is less than 10 ml/min. If the alarm message remains on the
screen, it indicates that the module does not leak.
Microstream: After 3 seconds, the alarm message "CO
2
Purging" is displayed on the screen. Block the gas
inlet for another 30 seconds. If the alarm message "
CO
2
Airway Occluded" is displayed, it indicates that the
module does not leak.
5.2.3.2 Accuracy Test
Tools required:
Steel gas cylinder with CO
2
(concentration range 3%~7%) and balance gas N
2
T-shape connector
Air tubing
Flowmeter
1. Connect the module and the monitor.
2. Wait until the CO
2
module warmup finishes. Check the airway for leakage and perform a leakage test as well to
make sure that the airway has no leakage.
3. Select Main Menu System Maintenance Module CO
2
Calibrate.
4. Connect the test system as follows:
Relief
valve
T-shape connector
Gas
cylinder
Tubing
Monitor
Flowmeter
5. Open the relief valve and adjust it until the flowmeter has a stable reading between 10 ml/min and 50 ml/min.
6. Check the real-time CO
2
value in the Calibrate menu and make sure the variation from the actual concentration is
within±0.3%.
5.2.3.3 Module Calibration
Tools required:
Steel gas cylinder with 6±0.05% CO
2
and balance gas N
2
T-shape connector
Air tubing
1. Make sure that the sidestream or microstream CO
2
module has been warmed up or started up.
2. Check the airway for leakage and perform a leakage test as well to make sure that the airway has no leakage.

Table of Contents

Related product manuals