13.1 Automatic test procedures
Appendix
5284 / 08/2023 enBENNING ST 755+ / ST 760+102
13 Appendix
13.1
Automatic test procedures
13.1.1 Automatic test procedures according to EN50678
(VDE0701) and EN50699 (VDE0702)
Testno. Name Individual tests
1 Devices of ClassI R
PE
(600mA), R
Insu‑1
(500V), functional test with
I
PE
(differential current measuring method)
2 Devices of ClassI with R
PE
10A R
PE
(10A), R
Insu‑1
(500V), functional test with I
PE
(differential current measuring method)
3 Devices of ClassI with I
Cont dir.
R
PE
(600mA), R
Insu‑1
(500V), I
Cont
(dir.), functional
test with I
PE
(diff.)
4 Devices of ClassI without R
PE
+
I
Cont dir.
Devices of ClassI without R
PE
, with I
Cont dir.
5 Devices of ClassI
heating<3.5kW
R
PE
(600mA), R
Insu‑1
(500V, R≥0.3MΩ),
functional test with I
PE
(diff.), for testing heating
devices with P<3.5kW
6 Devices of ClassI with I
PE
alt. R
PE
(600mA), R
Insu‑1
(500V), I
PE
(alternative
leakage current measuring method)
7 Devices of ClassI with R
PE
10A
+ I
PE
alt.
R
PE
(10A), R
Insu‑1
(500V), I
PE
(alternative leakage
current measuring method), for testing extension
cables up to 5m or tools with 300Hz such as
power screwdrivers, grinding tools
8 Devices of ClassI with R
Insu
250V
R
PE
(600mA), R
Insu‑1
(250V), functional test with
I
PE
(diff.), for testing devices with overvoltage
arresters
9 Devices of ClassI without R
Insu
R
PE
(600mA), functional test with I
PE
(diff.), test
procedure without R
Insu
only with factual basis
and justification
10 Devices of ClassI without R
Insu
+
I
Cont dir.
R
PE
(600mA), I
Cont
(dir.), functional test with I
PE
(diff.), test procedure without R
Insu
only with
factual basis and justification
11 Devices of ClassI with Uout. R
PE
(600mA), R
Insu‑1
(500V), R
Insu‑2
(500V),
R
Insu‑3
(500V), I
Cont
(dir.), functional test with I
PE
(diff.), U
a
at the secondary output (max. 25V),
e.g. for testing mains supply units and chargers
12 Devices of ClassI with I
PE
clamp R
PE
(600mA) and R
Insu‑1
(500V) with
BENNINGCM9‑1/ CM9‑2 and adapter,
itemno.: 044127/ 044128, I
PE
(clamp
measurement)
13 Devices of ClassI R
PE
10A with
I
PE
clamp
R
PE
(10A) and R
Insu‑1
(500V) with
BENNINGCM9‑1/ CM9‑2 and adapter,
itemno.: 044127/ 044128, I
PE
(clamp
measurement)
14 Shock-proof cable with probe
R
Insu
500V
R
PE
(600mA), R
Insu‑1
(500V), cable (line length
5m/ line cross-section 1.5mm²/ number of
conductors=3/ resistance value per line 0.3Ω,
for line tests of shock-proof cables