7.3 User administration
Functions
5284 / 08/2023 enBENNING ST 755+ / ST 760+36
7.3 User administration
The "User administration" menu is intended to manage the following user data and access
rights [}page64]:
User name
User role (admin, user)
Company
Postal code
Place
Password
The created users can be selected when logging on to the device.
User roles
Users with guest status (user role: user) do not have access to the user administration and
expert settings of the device after login. You do not need any password to log in.
Users with admin status (user role: admin) are granted access to all functions of the device
after login. You have to enter a password to log in.
7.4 Expert settings
The "Expert Settings" menu is intended to manage the following data:
Customer-specific limits
Customer-specific test procedures
Customer-specific visual inspection
Templates (customer-specific test questions and test procedures can be managed via the
"Templates" menu item)
Reset to factory settings
Updating the GUI(Graphical User Interface) and the firmware
The expert settings can be changed only by users with admin status.
7.4.1 Customer-specific limits
The "Customer-specific limits" [}page60]menu is intended to manage limits for manual tests
and customer-specific test procedures.
The following limits can be modified:
R
PE
(protective conductor resistance, line length, line cross-section)
R
Insu
(insulating resistance, testing voltage)
Cable (cable specifications)
U
a
(max. output voltage and open-circuit voltage of welding circuit)
PRCD (tripping current, tripping time, contact voltage)
I
Cont
(contact current values, differential current values)
I
Leak
(protective conductor current values, leakage current values, patient leakage current
values)
Test times (definition of all test times for individual tests)
The preset limits correspond to the test specifications that have been applicable at the time of
printing. Administration and definition of the customer-specific limits shall be at the discretion of
the qualified electrician carrying out the work.