8.1 Setting the system data
Configuration
5284 / 08/2023 enBENNING ST 755+ / ST 760+56
8 Configuration
8.1
Setting the system data
The "System settings”menu is intended to edit the basic settings of the device.
Menu
“Settings> Systemsettings> Systemdata“
Setting options
Parameter Description
Auto‑ID is enabled.
ON= The ID is automatically incremented [}page33].
Auto-ID start Start value that is incremented when a new customer is created or
a customer has no devices.
If test samples are already entered, the next free ID will be
determined automatically.
RPE test repetition ON= During the protective conductor resistance measurement, the
device automatically detects whether a measuring point is
contacted with the test probe.
If the measured value is "OK", the device responds with an acoustic
signal and the next measuring point can be contacted.
If the connection is not OK, there will be two subsequent acoustic
signals.
Test repetition ON= The device asks whether another measuring point shall be
contacted with the test probe.
The query applies to: R
PE
(only if the R
PE
test repetition is switched
off), R
Insu-2
, R
Insu-3
, R
Insu-4
, I
Cont
, I
Leak
, I
PLeak
, I
Cont weld.
Notification sound ON= Switches on the acoustic signal of the device.
Short-circuit test ON= Enables the short-circuit test before starting a measurement.
OFF= Disables the short-circuit test before starting a
measurement.
ITnetwork ON= Allows the device to be used in an ITnetwork.
Mains pole reversal OFF= Disables mains pole reversal for leakage current
measurements.
Standard test procedures ON= Enables the standard test procedures.
OFF= Hides the standard test procedures when the custom test
procedures are enabled.
Custom test procedures ON= Enables the custom test procedures.
OFF= Hides the custom test procedures when the standard test
procedures are enabled.
Polarity testing extension
cable
ON= Enables the polarity reversal test for extension cables.