EasyManua.ls Logo

Benning ST 755+ - Operating the Smart Menu; Table 13 Proceeding with the Test Report; Table 14 Smart Menu - Main Menu

Benning ST 755+
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General operation
6.7 Operating the Smartmenu
5284 / 08/2023 en BENNING ST 755+ / ST 760+ 33
Then, you can proceed as follows with the test reports:
Button Description
Print If a log printer is available, the "Test result" test report can be printed
directly.
Discard Deletes the "Test result" test report and finishes the test procedure.
Save Saves the "Test result" test report in the database. You can read it
out via your PC [}page35].
Table13: Proceeding with the test report
6.7 Operating the Smartmenu
In the automatic testing function, you can find the “Smartmenu” [}page29]. In the Smartmenu,
use the “Device templates” (test sample templates) and the “Auto-ID” function to quickly create
a new test sample with the subsequentID. This speeds up the creation of new test samples.
Button Description
Device template Here, you can select from previously created test sample templates
[}page63].
Device template–
Quick selection
button
Tap the quick selection button to create test samples and edit them as
required. These test samples will be assigned to the selected database
and the customer. If the “Auto-ID” function is enabled, the next free ID
will be assigned to the test sample. If the “Auto-ID” [}page56]
function is not enabled, the “Device” screen will be displayed to enter
the ID (test sample properties).
*Example of a quick selection button for a test sample of type "PRCD-
AC"
Device data
Quick access to the device data [}page58].
Favourites Quick access to menus and manual measurements.
Auto‑ID Intended to specify the next free auto-ID. This might be necessary, for
example, if a barcode roll with a new, consecutive (n+1) identification
number shall be used.
Standards Select the test standard. (Filter function for displayed test sample
templates)
Page selection "Scrolls" to the next page.
Table14: Smartmenu– Main menu
Procedure– Creating a test sample
1. Select a database and a customer.
2. Create a test sample (device). To do this, go to the "Device templates" menu, tap the
"Standards" button (note the page selection) and select the corresponding test sample
template [}page37] (quick selection button).
If the "Auto-ID" [}page56] function is not enabled, enter the desired ID.

Table of Contents

Related product manuals