53
of the analizer.
After this test press New Rotor
for continue working.
DDC112/Photometry
D: Choose Mode
- DDC112 internal test mode
- DDC112 Photometry Mode
- Stop
+: Integration Time +0.5ms
-: Integration Time -0.5ms
Notes: Only works in Service Mode
This tests performs continuous
readings with the DDC112.
Remember stop the test for
continue working.
Caution: Dont't abuse of this functions while the
analizer is running.
CONNECTED WITH PC!
Rx s=0
Stand by mode!
Send Status
Rx s=1
Inicio modo servicio!
Hardware Initialization
Programming FPGA XCS30TQ144
- Clearing FPGA program memory: OK!
- Programing FPGA: OK
FPGA XCS30TQ144 is programmed
Initialization of level detection system
- Generating Sensitivity Map:Ok
- Programming X9279: .OK
Sensitivity is programmed
Initializating Motors
- Axis Z in HOME
- Axis Y in HOME
- Axis X in HOME
- ROTOR in HOME
- FILTERS WHEEL in HOME
- CERAMIC PUMP in HOME
Motors Initializated
Optics Initialization
- Filter correction=4
- Rotor correction=1113 Rotor Lect. correction=-4
- Used wells:28
- Check DDC112.OK (FPGA 8Mhz Clock: OFF)
- DDC112 test: OK! (Result => 0045140)
Optics Initializated
Generating Pattern: 1 1 1 1 1 1 1 1 1 0 0 0 0 0 0 0 0 0 0 0
Generating Gate: 0 0 0 1 1 1 0 0 0 0 0 0 1 1 1 1 1 0 0 0
OK
Z Axis Initialization
- Axis Z in HOME
Z Axis initializated
Programming the encoder: 0000| OK
Z Axis Initialization
- Axis Z in HOME
Z Axis initializated
Auto-Adjustment of the probe
- Positioning ARM
- Detecting X side: OK. The correction is X=Xcal+0