Service manual
52
- Rotor: 88 Steps
- Filter Wheel: 1771 Steps
- Ceramic Pump: 1418 Steps
- Washing Station Pump: 903 Cycles
- Washing Station Valve: 903 Cycles
- Ceramic Pump Valve: 1341 Cycles
- Lamp: 661 Minutes
A25 Statistics
- Biochemistry Tests: 0
- Turbidimetry Tests: 171
- Biochemistry Bireagent Tests: 0
- Turbidimetry Bireagent Tests: 0
- Predilutions: 0
- Initial/Final Washings: 0
- Washing Solution Washings: 0
- System Liquid Washings: 50
- New Rotor: 1
- Bireagent Contaminations Solved: 0
Detecting Main Board version (CIIM00006-XX)
Ok. Main Board version is CIIM00006-02/03
Configuration Load from X9279: STT= 3 RD=0 TD=1 GD=0
Setting racks layout
Tray Ref. X=1349 => Distance from tray reference to tray corner X=1082
Tray Ref. Y=335 => Distance from tray reference to tray corner Y=432
Absolute position of tray corner X=267
Absolute position of tray corner Y=-97
Generating Zmax Map:Ok
CPU settings: MDCR=c4;ABWCR=0;ASTCR=ff
BioSystems A25
Hello World
A25 MAGIC KEYS
H: Help
R: Rotor Temperature
P: Probe Temperature
S: Level Scales
A: Last A25 Stress Results
L: Actual Sensitivity of Level Detection
N: Enable Level Detection Debug
K: Power Supply On
Buzzer Control
B: Buzzer On
b: Buzzer Off
Encoder
E: Generate Encoder Error
I: Enable Encoder IRQ
Rotor Reading
1: Choose Filter +
2: Choose Filter -
9: Start Rotor Readings
Notes: Use only in Service Mode
after a Base Line Test.
Rotor Read
1: Choose Filter +
2: Choose Filter -
User Mode Test
G: Test
Notes: Use only after a Worklist in
Stand By.
This tests dumps all the preparations
parameters received and the photometric
readings. Finally performs a general test