SK77981-14
184 | R-net Technical Manual | Chapter 10 Diagnostics
Curtiss-Wright | PG Drives Technology
3.1 CENTER JOYSTICK
The most common cause of this trip is if the joystick is deflected away from center before and during the time the control
system is switched on. The joystick displaced screen will be displayed for 5 seconds, if the joystick is not released within that
time then a trip is registered.
Ensure that the joystick is centered and power-up the control system.
If the trip is still present then the joystick or Joystick Module may be defective.
CAUTION:
Depending on the Wheelchair Manufacturer’s service agreement, the joystick may be replaced as a separate part. Otherwise
refer to section 5.
3.2 LOW BATTERY
This occurs when the control system detects that the battery voltage has fallen below 16V.
Check the condition of the batteries and the connections to the control system.
If the trip is still present after the batteries and connections have been checked, then the Power Module may be defective.
Refer to section 5.
3.3 HIGH BATTERY VOLTAGE
This occurs when the control system detects that the battery voltage has risen above 35V. The most common reasons for this
are overcharging of the battery or bad connections between the control system and the batteries.
Check the condition of the batteries and the connections to the control system.
If the trip is still present after the batteries and connections have been checked, then the Power Module may be defective.
Refer to Section 5.
3.4 BRAKE ERROR
This occurs when the control system detects a problem in the solenoid brakes or the connections to them.
1505 - M1 Brake Error - Refer to section 2.3 for connection details
1506 - M2 Brake Error - Refer to section 2.3 for connection details
Check the solenoid brakes, cables and connections to the control system.
If the trip is still present after the above checks have been made, then the Power Module may be defective. Refer to Section 5.
3.5 MOTOR ERROR
This occurs when the control system detects that a motor has become disconnected.
3B00 - M1 Motor Error - Refer to section 2.3 for connection details
3C00 - M2 Motor Error - Refer to section 2.3 for connection details
Check the motors, cables and connections to the control system.
If the trip is still present after the above checks have been made, then the Power Module may be defective. Refer to Section 6.