EasyManua.ls Logo

Endress+Hauser Proline Prosonic Flow P 500 - Page 129

Endress+Hauser Proline Prosonic Flow P 500
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Proline Prosonic Flow P 500 HART Commissioning
Endress+Hauser 129
Parameter Prerequisite Description Selection Factory setting
Process variable unit 1 to n A process variable is selected
in the Assign process variable
parameter (→  128) of the
Totalizer 1 to n submenu.
Select the unit for the process
variable of the totalizer.
g
*
kg
*
t
*
oz
*
lb
*
STon
*
cm³
*
dm³
*
*
ml
*
l
*
hl
*
Ml Mega
*
af
*
ft³
*
Mft³
*
Mft³
*
fl oz (us)
*
gal (us)
*
kgal (us)
*
Mgal (us)
*
bbl (us;liq.)
*
bbl (us;beer)
*
bbl (us;oil)
*
bbl (us;tank)
*
gal (imp)
*
Mgal (imp)
*
bbl (imp;beer)
*
bbl (imp;oil)
*
MSft³
*
None
*
Depends on country:
ft³
Totalizer 1 to n operation mode A process variable is selected
in the Assign process variable
parameter (→  128) of the
Totalizer 1 to n submenu.
Select totalizer operation
mode, e.g. only totalize
forward flow or only totalize
reverse flow.
Net
Forward
Reverse
Net flow total
Totalizer 1 to n failure behavior A process variable is selected
in the Assign process variable
parameter (→  128) of the
Totalizer 1 to n submenu.
Select totalizer behavior in the
event of a device alarm.
Hold
Continue
Last valid value +
continue
Stop
* Visibility depends on order options or device settings

Table of Contents

Other manuals for Endress+Hauser Proline Prosonic Flow P 500

Related product manuals