Commissioning Proline Prosonic Flow P 500 HART
132 Endress+Hauser
Parameter Prerequisite Description Selection / User
entry
Factory setting
100% bargraph value 3 A selection was made in the
Value 3 display parameter.
Enter 100 % value for bar
graph display.
Signed floating-point
number
0
Decimal places 3 A measured value is specified
in the Value 3 display
parameter.
Select the number of decimal
places for the display value.
• x
• x.x
• x.xx
• x.xxx
• x.xxxx
x.xx
Value 4 display A local display is provided. Select the measured value that
is shown on the local display.
For the picklist, see
Value 1 display
parameter
(→ 122)
None
Decimal places 4 A measured value is specified
in the Value 4 display
parameter.
Select the number of decimal
places for the display value.
• x
• x.x
• x.xx
• x.xxx
• x.xxxx
x.xx
Display language A local display is provided. Set display language. • English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык
(Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• tiếng Việt
(Vietnamese)
• čeština (Czech)
English
(alternatively, the
ordered language is
preset in the device)
Display interval A local display is provided. Set time measured values are
shown on display if display
alternates between values.
1 to 10 s 5 s
Display damping A local display is provided. Set display reaction time to
fluctuations in the measured
value.
0.0 to 999.9 s 0.0 s
Header A local display is provided. Select header contents on local
display.
• Device tag
• Free text
Device tag
Header text The Free text option is
selected in the Header
parameter.
Enter display header text. Max. 12 characters,
such as letters,
numbers or special
characters (e.g. @,
%, /)
------------
Separator A local display is provided. Select decimal separator for
displaying numerical values.
• . (point)
• , (comma)
. (point)
Backlight One of the following
conditions is met:
• Order code for "Display;
operation", option F "4-line,
illum.; touch control"
• Order code for "Display;
operation", option G "4-line,
illum.; touch control
+WLAN"
Switch the local display
backlight on and off.
• Disable
• Enable
Enable
* Visibility depends on order options or device settings