Mounting Proline Prosonic Flow P 500 HART
36 Endress+Hauser
M6
50 mm
(2 in)
1
2
3
1
13 mm
A0043375
 28 Holder with welded bolts
1 Welding seam
2 Locking nut
3 Hole diameter max. 8.7 mm (0.34 in)
Installing sensor – small nominal diameters DN 15 to 65 (½ to 2½")
Requirements
The installation distance is known. →  30
Sensor holder is pre-assembled.
Material
The following material is required for mounting:
Sensor incl. adapter cable
Sensor cable for connection to the transmitter
Coupling medium (coupling pad or coupling gel) for an acoustic connection between the
sensor and pipe
Procedure:
1. Set the distance between the sensors to the value determined for the sensor distance.
Press the movable sensor down slightly to move it.
b
a
A0043376
 29 Distance between sensors as per the installation distance →  30
a Sensor distance (back of sensor must touch the surface)
b Sensor contact surfaces
2. Stick the coupling pad under the sensor onto the measuring pipe. Alternatively, coat
the contact surfaces of the sensor (b) evenly with coupling gel (approx.
0.5 to 1 mm (0.02 to 0.04 in)).