Proline Prosonic Flow P 500 HART Mounting
Endress+Hauser 39
Procedure with a tape measure:
1. Use a tape measure to determine the pipe diameter d.
2. Mount the opposite mounting bolt at d/2 from the front mounting bolt. The distance
must be d/2 = d'/2 on both sides.
3. Check distance B.
d/2 = d‘/2
B
d/2
d‘/2
A0052445
 34 Positioning the strapping bands and mounting bolts with a tape measure (steps 2 to 4)
Fastening the sensors:
1. Fit the sensor holders over the individual mounting bolts and tighten securely with
the locking nut.
13 mm
A0043381
 35 Mounting the sensor holders
2. Stick the coupling pad under the sensor →  217. Alternatively, coat the contact
surfaces of the sensor evenly with coupling gel (approx. 1 mm (0.04 in)). In doing so,
start from the groove through the center to the opposite edge.
A0043382
 36 Coating the contact surfaces of the sensor with coupling gel (if there is no coupling pad)
3. Insert the sensor into the sensor holder.