Operation options Proline Prosonic Flow P 500 HART
84 Endress+Hauser
Via WLAN interface
The optional WLAN interface is available on the following device version:
Order code for "Display; operation", option G "4-line, illuminated; touch control + WLAN"
1
2
3
4
3
4
5 6 7
A0041325
1 Transmitter with integrated WLAN antenna
2 Transmitter with external WLAN antenna
3 LED lit constantly: WLAN reception is enabled on measuring device
4 LED flashing: WLAN connection established between operating unit and measuring device
5 Computer with WLAN interface and web browser (e.g. Microsoft Internet Explorer, Microsoft Edge) for
accessing the integrated device web server or with operating tool (e.g. FieldCare, DeviceCare)
6 Mobile handheld terminal with WLAN interface and web browser (e.g. Microsoft Internet Explorer, Microsoft
Edge) for accessing the integrated device web server or operating tool (e.g. FieldCare, DeviceCare)
7 Smart phone or tablet (e.g. Field Xpert SMT70)
Function WLAN: IEEE 802.11 b/g (2.4 GHz)
Encryption WPA2-PSK AES-128 (in accordance with IEEE 802.11i)
Configurable WLAN channels 1 to 11
Degree of protection IP67
Available antennas Internal antenna
External antenna (optional)
In the event of poor transmission/reception conditions at the place of
installation.
Only 1 antenna is active at any one time!
Range Internal antenna: typically 10 m (32 ft)
External antenna: typically 50 m (164 ft)
Materials (external antenna) Antenna: ASA plastic (acrylonitrile styrene acrylate) and nickel-plated
brass
Adapter: Stainless steel and nickel-plated brass
Cable: Polyethylene
Plug: Nickel-plated brass