EasyManua.ls Logo

Fluke 9500B - Page 110

Fluke 9500B
297 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.2-2 Section 5: 9500B Procedure Mode - Safety and General Notes
Final Width = 215mm
Descriptions assume 9500B/1100
5.2.4 Printing Setup
The results of adjustment and verification
operations on UUTs can be printed on one of
two forms of certificate. A suitable printer
must be connected and switched on-line, and
the required certificate style, format and data
must be entered into the 9500B memory. Then
with the correct printer type enabled, the 9500B
internal program will generate the required
certificate.
5.2.4.1 Printer Type
The printer to be used should be capable of
printing 120 characters per line, and must be
able to print the Code Page 437 character set.
Most printers compatible with Epson FX,
Canon Bubble-Jet or Hewlett-Packard Desk-
Jet are suitable. The printer is connected to the
25-way D-type port on the 9500B rear panel.
5.2.4.2 Certificate Formatting and
Data Presentation
Config mode is used to select the style of
certificate to be printed, and to set the format of
page length, headers, footers, etc. In addition,
such certificate entries as laboratory
identification, temperature and humidity can
be added. Details of these elements of Config
mode can be found in Section 3, paras 3.4.3.
Note: If the Procedure Card used to run the
procedure was originally written for
the Model 9100 Option 250 or 600,
then the procedure and certificate will
incorporate the relevant 9100 (not
9500B) uncertainties.
5.2.4.3 Enable Printing
Printing must be enabled, using the Config
mode screen keys 'PRINTER' and the
appropriate selection of printer type (refer to
Section 3, paras 3.4.3.5).

Table of Contents

Other manuals for Fluke 9500B