EasyManua.ls Logo

Fluke DTX Series - About LinkWare and LinkWare Stats Software

Fluke DTX Series
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DTX Series CableAnalyzer
Technical Reference Handbook
2-22
For example, the following start and stop IDs could be
used for testing the cabling in two rooms where each
room has three cable drops:
Start ID: ROOM A DROP#1
Stop ID: ROOM B DROP#3
These IDs produce the following ID list:
ROOM A DROP#1
ROOM A DROP#2
ROOM A DROP#3
ROOM B DROP#1
ROOM B DROP#2
ROOM B DROP#3
The steps below guide you through creating a sequential
ID list for the following scenario:
You will test 12 cables in two patch panels: cables 1
through 6 in panel A and cables 1 through 6 in panel
B.
Both panels are located in telecommunications closet
A on the third floor of the building.
Your IDs will follow the ANSI/TIA/EIA-606-A standard for
horizontal links. See “About ANSI/TIA/EIA-606-A Cable
IDs” on page 2-24 for details.
1 Turn the rotary switch to SETUP.
2 Use
D to highlight Instrument Settings, then
press
H.
3 Press
H to select Cable ID Source.
4 Use
D to highlight Auto Sequence, then press
H.
5 Use
D to highlight Template, then press H.
6 Use
A D to highlight Horizontal, then
press
H.
7 On the Auto Sequence screen, press
J Default to
transfer the default pattern into the START ID and
STOP ID.
8 Use A to highlight Start ID; then press H.
9 Use
JKL, BCAD, and H to
change the default ID to 03A-A01.
Press
N when you are done.
10 Use D to highlight Stop ID; then press H.
11 Use
JKL, BCAD, and H to
change the default ID to 03A-B06.
Press
N when you are done.

Table of Contents

Other manuals for Fluke DTX Series

Related product manuals