Tutorials on Setup and Test Procedure
Using the Auto Increment and Sequential Cable ID Features 2
2-23
12 Press L Sample List. You should see a list of 12
sequential fiber IDs: 03A-A01 through 03A-B06. If the
tester beeps instead of showing the list, see “Solving
Problems with ID Lists” on page 2-25.
13 Press
I to leave the sample list.
If the tester shows a error message, check your Start
and Stop IDs for the problems listed below.
14 Press
N when you are done setting up the list. The
list is saved in the tester’s internal memory.
If the ID sequence is invalid, check the following:
• Verify that the types of characters in each position
match between the start and stop IDs. For example,
using the letter “O” as the third character in the Start
ID and the number “0” as the third character in the
Stop ID is not allowed.
• Verify that you are not using the characters / # * - . , :
[ ] space or accented characters as incrementing
characters. You may use these characters in IDs, but
they must match between the Start and Stop IDs.
• Verify that the Start and Stop IDs have the same
numbers of characters.
• Verify that the Stop ID is not sequentially greater
than the Start ID. For example, using 25 as the Start
ID and 10 as the Stop ID is not allowed.
• Verify that the Start and Stop IDs do not generate a
sequence with more than 3000 IDs.
To use an ID from the list:
1 Run a test; then press
N.
2 Select an ID from the ID list; then press
N again.