Infratecâ„¢ 1241 Grain Analyzer
4:32 Service Manual 1001 5015 / Rev. 4
33538 0x8302 TWM operation mode switch.
33539 0x8303 TWM EEPROM error Problems to read/write the data
in the PROM
33540 0x8304 TWM EEPROM error Slope data outside limits. try to
re calibrate the balance.
34050 0x8502 Lamp On/Off error.
34051 0x8503 Set display On/Off error.
34052 0x8504 Set back light On/Off error.
34053 0x8505 Set contrast On/Off error.
1 0x0001 Instrument communication er-
ror.
2 0x0002 Detector ADC error There is no signal from the De-
tector Board.
4 0x0004 Monochromator grating error. See
4.11.1
22 0x0016 Measuring Unit TWM data
range error (invalid data)
See
4.11.2
32 0x0020 Cell blocked error Sample Cell is dirty. Analysis
cancelled. Empty and clean the
Cell using the flush function on
the Task Bar. More informa
-
tion is to be found in User
Manual
See
4.11.3
64 0x0040 More sample required. More
sample is needed to fill up the
Sample Cell.
Displayed if the Application
Model allows refill.
128 0x0080 Path Length error. See
4.11.4
512 0x0200 Speckle function error. See
4.11.5
1024 0x0400 Warning! Drawer not in place. If the drawer is in place and
you still get this warning, then
check the Drawer sensor.
2048 0x0800 TWM shutter error.
4096 0x1000 TWM wiper error.
8192 0x2000 TWM not full.
16384 0x4000 TWM load cell error. See
4.11.6
32768 0x8000 TWM EEPROM error.
Failed to read Hopper tempera-
ture sensor.
4242 0x1092 User abort
SL
Error
HL
Error
Message to the User (What
possibly can be done before
contacting Service)
Comments Note