3.4 Installation of Sample Transport Module
Software Options
In the Service Menu there is an option in “3 Manual Tests” called “7 STM” which
offers three options.
1. Position settings
2. Service
3. Exercise
Position settings can be used to run the cuvette holder up and down manually. It can
also be used to step between the sub sample positions. It is easy to use and stops when
it comes to the end position.
Service, displays the position sensors value and gives an option of running the motor
Up/ Down/ Off.
Note: It will not stop in the end position so be very careful.
Exercise, will only continuously run the cuvette holder up and down, no real use for
this option.