Infratec™ 1241 Grain Analyzer
5:74 Service Manual 1001 5015 / Rev. 4
5.7.2 TWM Balance Calibration
For the calibration you need the Test Weight Control Kit, p/n 10011779.
The content of the kit is:
2 pcs of crowned 200 g weight
1 pcs weighing bridge with fastening nut.
Fig. 5:75
1. Log in on the Service menu. The password is 3677. A hint to remember this
password is to type "FOSS" on you mobile phones keyboard.
2. Select Manual Tests/ Test weights/ Calibration/ Mass calibration (3.6.2.2).
3. Basically just follow the instructions given on the display.
Fig. 5:76