Infratecâ„¢ 1241 Grain Analyzer
Service Manual 1001 5015 / Rev. 4 5:65
5.5 Sample Transport Module
5.5.1 Replacement of Cuvette Glass Kit
If the cuvette glass is broken or needs to be exchanged for other reasons, follow the
procedure below:
1. Remove glass and adhesive remains from the cuvette.
2. Clean the cuvette and wipe it dry.
3. Tear of the protective tape from the adhesive and glue the new glass to the
cuvette.
Applicable to:
1001 2122 Cuvette bottom part
1001 2123 Cuvette top part 6 mm
1001 2124 Cuvette top part 10 mm
1001 2125 Cuvette top part 18 mm
1001 2126 Cuvette top part 25 mm
1001 2127 Cuvette top part 29 mm
1001 3213 Top loaded cuvette 6/10/18 mm