EasyManuals Logo

Foss Infratec 1241 Service Manual

Foss Infratec 1241
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #70 background imageLoading...
Page #70 background image
Infratec™ 1241 Grain Analyzer
4:2 Service Manual 1001 5015 / Rev. 4
4.1.1 Poor Results
The points listed below can be considered as a check list for trouble shooting “poor”
results. The reasons for “poor” results can be divided into:
User problems
Poor calibrations
Outliers
Poor reference data
Instrument problems (Time for cleaning? see User Manual)
4.1.2 User Problems
Are they using the right calibration?
Has there been an adjustment to the slope or intercept in the calibration?
Are they sure that the moisture basis of the Infratec results is the same as that of
the reference results?
Do they use adjustment to dry content or constant moisture content?
How many sub-samples are used?
4.1.3 Poor Calibrations
Was the calibration developed for this type of sample?
Was there sufficient data to develop the calibration?
Was the calibration stabilised for instrument and temperature differences?
4.1.4 Outliers
Results with outliers should always be confirmed by a second analysis.
A sample may be classified as an outlier for a number of reasons:
The constituent concentration is outside of the calibration range.
The sample is of a variety for which the NIT spectrum differs greatly from the
samples in the calibration set.
The product is analysed with wrong AM.
There is too much foreign material in the sample.
Movements of the sample during the scan.
Too high absorbance.
Packing of sample not good
For each constituent, an outlier will be displayed as a letter (A-E) and a number
(1-5). The letter indicates what kind of outlier it is, as described below, and the
number describes the severity (the higher the number, the stronger the outlier).
A-outlier (residual) and B-outlier (leverage)
An outlier denoted by the letters A and B indicate how closely the NIT spectra of the
sample correlate with the spectra in the calibration set.
C-outlier (sub sample deviation)
An outlier denoted by the letter C indicates that the standard deviation between the
sub predicted values from the individual sub samples is above the set limit.
D-outlier (out of range)
An outlier denoted by the letter D indicates that the low or high limit for the constit-
uent has been exceeded by the predicted value of at least one of the sub samples. This
means that it is possible to get a D-outlier even if the average result on the display is
within the calibration range.

Table of Contents

Other manuals for Foss Infratec 1241

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Foss Infratec 1241 and is the answer not in the manual?

Foss Infratec 1241 Specifications

General IconGeneral
BrandFoss
ModelInfratec 1241
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals