Infratec™ 1241 Grain Analyzer
Service Manual 1001 5015 / Rev. 4 5:77
5.7.3 TWM balance Control
1. Select Manual Tests/ Test weights/ Calibration/ Balance Control (3.6.2.1).
Fig. 5:82
2. Follow the instructions on the display. After completion the Software will
answer the question "Balance OK?" with either Yes or No. If the Expected and
Calculated value differs more than +/- 0.2 the answer is No.
In this case check the following;
That no vibrations are present.
That no cables or other objects are touch the exterior of the balance.
If service was done on internal parts make sure nothing is touching the balance.