Infratec™ 1241 Grain Analyzer
5:38 Service Manual 1001 5015 / Rev. 4
5.3.3 Replacement of Upper Solenoid
Note: The replacement is carried out with the Measuring Unit
standing apart.
1. Disconnect the Upper Solenoid Cable from the Measuring Unit Board.
2. Remove the two screws holding the Upper Solenoid.
3. Exchange the Upper Solenoid and assemble in reverse order.