Infratecâ„¢ 1241 Grain Analyzer
Service Manual 1001 5015 / Rev. 4 4:21
4.4 Automatic Tests
From here all of the tests carried out in the Self-test routine can be executed individ-
ually. The test results are presented in the same way as described in section Self-test,
with an option to save to disk.
Fig. 4:7 Automatic Tests
4.4.1 Detector
Fig. 4:8 Detector Tests
ADC No Missing Codes
Identical to the "ADC No Missing Codes" test carried out in the Self-test routine.
Offset with Empty Cell
Identical to the "Detector Offset" test carried out in the Self-test routine.