Infratecâ„¢ 1241 Grain Analyzer
4:44 Service Manual 1001 5015 / Rev. 4
4.12.6 Error Word 2
"Sub-error", presents a more detailed information about what went wrong when the
error occurred.
Position in instrument Hardware Error Code in bold:
E.g. "20 - 304/1ef3/201/20/0/0/0/450".
Id (hex) Specification Note Source
0001 SystemCommandFailed Unknown command All modules
0002 ISWIllegalCommand Illegal command from ISW All modules
0004 ISWUnknownCommand Unknown command DSP
0008 ISWChecksumError Checksum error DSP
0010 ISWCommandHeadError Command header error DSP
0020 ISWCommandTailError Command tail error DSP
0040 MonochromatorTimeOut
Monochromator comm. timed
out
Monoch.
0080 DetectorADCFailed ADC failed Detector
0100 MUTimeOut MU communication timed out MU
0200 TWMTimeOut TWM communication timed out TWM
0400 GratingStartupFailed Grating startup procedure failed DSP
0800 Timeout
I
2
C timeout
All Modules
1000
I
2
C Locked I
2
C bus line locked
I
2
C -
modules
2000
I
2
C NoAck No I
2
C acknowledge from slave
I
2
C -
modules
4000
I
2
C ChecksumError I
2
C checksum error
I
2
C -
modules
8000 CommandFailed Communication failed DSP