Infratec™ 1241 Grain Analyzer
5:64 Service Manual 1001 5015 / Rev. 4
3. If the level sensor is in a wrong position - loosen the screws 1. Adjust the level
sensor to correct position and tighten the screws 1.
Fig. 5:66 Adjusting level sensor
4. Assemble in reverse order.
5. Start up the instrument and let it go through the Startup Test.
6. Perform a TWM balance calibration, see 5.7.2 TWM Balance Calibration on
page 5:74.
1 Screws