Infratecâ„¢ 1241 Grain Analyzer
Service Manual 1001 5015 / Rev. 4 4:49
HV/12416305/TWM/SerialNumber=String,"21436587"
HV/12416305/Eid=String,"64e7780e00c0"
Software Log (xxxxxxxx.SL)
Logs the messages shown on the display to the operator.
Example:
;0019,0255,0019,1000
S L / 1 2 4 1 5 2 3 8 / 2 0 0 7 - 1 0 - 1 6 2 0 : 0 4 : 4 1 / N a m e = S t r i n g , " E r r o r \ t S e l f t e s t s t o p p e d . "
SL/12415238/2007-10-18 13:34:09/Name=String, "Error\t Sample in the cell\t
Empty and clean the cell using the\t flush function on the taskbar\t More information
in the user manual. "
SL/12415238/2007-10-18 13:36:44/Name=String, "Error\t Analysis Failed\t Error
in Measuring Unit\t Please note Code 128 (030d)\t for support "
SL/12415238/2007-10-18 16:10:36/Name=String, "Error\t Sample in the cell\t
Empty and clean the cell using the\t flush function on the taskbar\t More information
in the user manual. "
SL/12415238/2008-02-08 14:19:56/Name=String, "Error\t Verifying Monoch
Checksum\t failed selftest\t Do you want to continue the self test ? "
SL/12415238/2008-02-08 14:27:00/Name=String, "Error\t Self test failed.\t Check
test results for further inform ation \t Se e the User Manual "
SL/12415238/2008-02-21 10:34:43/Name=String, "Error\t Verifying Monoch
Checksum\t failed selftest\t Do you want to continue the self test ? "
SL/12415238/2008-02-21 10:35:30/Name=String, "Error\t Self test failed.\t Check
test results for further inform ation \t Se e the User Manual "
SL/12415238/2008-05-27 15:24:47/Name=String, "Error\t Manual Scan Failed.\t
E r r o r i n M e a s u r i n g U n i t \ t P l e a s e n o t e C o d e 3 2 7 6 8 ( 0 3 0 e ) \ t f o r s u p p o r t "
Hardware Log (xxxxxxxx.HL)
Logs the command and response between the DSP and hardware I2C devices. This
file is by far the most useful during troubleshooting.
Example:
;0053,0255,0053,0100
HL/12415238/2007-10-17 10:29:42/Name=String, "40 - 30e/1fc3/201/40/0/0/0/
501"
HL/12415238/2007-10-18 13:34:02/Name=String, "20 - 30c/1fc3/201/20/0/0/0/
501"
HL/12415238/2007-10-18 13:34:06/Name=String, "20 - 30c/1fc3/201/20/0/0/0/
501"
HL/12415238/2007-10-18 13:34:09/Name=String, "200 - 30d/1fc3/201/200/0/0/0/
501"

Table of Contents

Other manuals for Foss Infratec 1241

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Foss Infratec 1241 and is the answer not in the manual?

Foss Infratec 1241 Specifications

General IconGeneral
BrandFoss
ModelInfratec 1241
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals