Infratec™ 1241 Grain Analyzer
4:56 Service Manual 1001 5015 / Rev. 4
SELFTEST/12417132/START STABILITY TEST/2011-01-18 15:43:18
SELFTEST/12417132/Check Lamp and Grating Motor = Instrument No Error
SELFTEST/12417132/Stability/Max=Number,47222 (>30000)
SELFTEST/12417132/Stability/Max=Number,47124 (>30000)
SELFTEST/12417132/Stability/Max=Number,47059 (>30000)
SELFTEST/12417132/Stability/Max=Number,47014 (>30000)
SELFTEST/12417132/Stability/Max=Number,46979 (>30000)
SELFTEST/12417132/Stability/Max=Number,46951 (>30000)
SELFTEST/12417132/Stability/Max=Number,46928 (>30000)
SELFTEST/12417132/Stability/Max=Number,46910 (>30000)
SELFTEST/12417132/Stability/Max=Number,46894 (>30000)
SELFTEST/12417132/Stability/Max=Number,46879 (>30000)
SELFTEST/12417132/Stability/Max=Number,46866 (>30000)
SELFTEST/12417132/Stability/Max=Number,46854 (>30000)
SELFTEST/12417132/Stability/Max=Number,46836 (>30000)
SELFTEST/12417132/Stability/Diff=Number,-170.34
SELFTEST/12417132/Stability/Max=Number,46829 (>30000)
SELFTEST/12417132/Stability/Diff=Number,-136.67
SELFTEST/12417132/Stability/Max=Number,46821 (>30000)
SELFTEST/12417132/Stability/Diff=Number,-114.33
SELFTEST/12417132/Stability/Max=Number,46814 (>30000)
SELFTEST/12417132/Stability/Diff=Number,-98.67
SELFTEST/12417132/END STABILITY TEST/2011-01-18 15:45:56
SELFTEST/12417132/START BG20FILTER TEST/2011-01-18 15:45:58
SELFTEST/12417132/Check Lamp and Grating Motor = Instrument No Error
SELFTEST/12417132/Monochromator/I/SerialNumber=String,E0113281
SELFTEST/12417132/Monochromator/I/Peak2Mean=Number,880.01
SELFTEST/12417132/Monochromator/I/Peak2Min=Number,880.01
SELFTEST/12417132/Monochromator/I/Peak2Max=Number,880.01
SELFTEST/12417132/Monochromator/I/Peak2Std=Number,0.00
SELFTEST/12417132/Monochromator/I/Peak2Diff=Number,-0.14 (>=-1.0 , <=1.0)
SELFTEST/12417132/END BG20FILTER TEST/2011-01-18 15:46:31
Self Test Log (xxxxxxxx.ST1 - xxxxxxxx.ST3)
The three latest self tests will be saved and named xxxxxxxx.ST1 - xxxxxxxx.ST3.
The files are automatically renamed so xxxxxxxx .ST1 is always the latest.