Infratecâ„¢ 1241 Grain Analyzer
4:24 Service Manual 1001 5015 / Rev. 4
The test runs the ports individually as chosen in the "I/0" menu. If any error occur
this is reported to screen and Error log.
Keyboard
The test is going through every key from upper left to down right by stating which
key to press. When the correct key is pressed the next key is stated.
If no key is detected within 10 seconds from last pressed key, the test aborts. If abort-
ed, this will be logged to Error log with the message on which key that should be
pressed when the sequence was aborted.
External Keyboard
The test shows which key on the external keyboard that is pressed at the moment. If
no key is detected within 10 seconds after the last pressed key,
the test aborts without any Error log message.
Line description including typical value Units Min. Max.
SelfTest/00000000/PCBoard/COMParallel=Bool, True bool
SelfTest/00000000/PCBoard/COM1Serial=Bool, True bool
SelfTest/00000000/PCBoard/COM2Serial=Bool, True bool