Revision C 250cx Series Maternal/Fetal Monitor A-5
2036946-001
Factory Defaults: Table of Defaults
Master
Alarm
Setup
Alarm Limits (mmHg)
Systolic
Diastolic
MAP
MHR/P
MSpO
2
High Low
160 90mmHg
90 50 mmHg
140 50 mmHg
120 50 bpm
100 95%
High Low
70-240 50-150
70-130 30-120
70-150 30-120
100-25 35-120
85-100 80-99
Alarm (kPa mode)
Systolic
Diastolic
MAP
MHR/P
High Low
21.3 12.0 kPa
12.0 6.7 kPa
18.7 6.7 kPa
120 50 bpm
High Low
9.3-32.0 6.7-20.0
9.3-17.3 4.0-16.0
9.3-20.0 4.0-16.
100-250 35-120
Volume 5 1-9
General
Setup
Play Song Off Off, Happy Birthday,
Brahms’ Lullaby, Rock-a-
Bye-Baby, All
Song Volume 5 0-9
Temp Done Volume 5 0-9
Brightness 9 0-9 (nine = brightest)
Paper Speed United States: 3 cm/min
International: 1 cm/min
1-3 cm/min
Date Set to current local date. Set to current local date.
Time Set to current local time.
Must manually change to
EST/EDT.
Set to current local time.
MSpO
2
Print Interval
(External Monitor)
5 min Off, 2, 5, 10, 15, 30, 60 min
FSpO
2
Print Interval
(External Monitor)
5 min Off, 2, 5, 10, 15, 30, 60 min
FSpO
2
Trace Off Off, On
Vital Signs
History
HX Interval Event 1, 5, 10, 15, 30, 60, Event
Summary of Factory Defaults
Setup
Screen
Field Description Factory Default Default Options Hospital Preference