Revision C 250cx Series Maternal/Fetal Monitor 13-5
2036946-001
Heartbeat Coincidence: Using the Heartbeat Coincidence Feature
Strip Chart Annotation
When heartbeat coincidence detection is enabled, the annotation HBC prints in the
center margin of the strip chart paper following the active FHR modes. (Refer to
“Simulated Heartbeat Coincidence Detection Trace,” on page 13-5.)
As soon as heartbeat coincidence is detected, two overlaid hearts print in the
upper portion of the top grid of the strip chart paper; the hearts print every 4.5 cm for
as long a coincidence is detected. Once coincidence is resolved, two side-by-side
hearts print once.
If you disconnect a transducer while coincidence is detected, the overlaid hearts
stop printing. In addition, the mode status line prints on the strip chart paper—
without the
HBC annotation—indicating the deactivated mode.
Simulated Heartbeat Coincidence Detection Trace
Coincidence detected.
Coincidence resolved.
Coincidence enabled.
This annotation prints
every 30 minutes along
with the modes.
Simulated trends.