EasyManua.ls Logo

GE Corometrics 250cx Series

GE Corometrics 250cx Series
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Revision C 250cx Series Maternal/Fetal Monitor 5-9
2036946-001
Fetal Heart Rate Monitoring: Fetal Heart Rate Alarms
100% Signal Loss
In cases where there is a complete absence of signal, the signal quality time-to-alarm
is 1.25 minutes.
100% Signal Loss Example
Intermittent Signal Loss
In the clinical environment, a partial loss of signal is seen more frequently than a
complete loss of signal. The time-to-alarm will vary related to the percentage of
signal loss. Figure below shows an example where there is 70% signal loss resulting
in a signal quality alarm after 5 minutes.
70% Signal Loss Example
Silencing an Audio Alarm
Press the Alarm Silence button to cancel the audio component of an alarm; the
visual indications remain until the alarm condition is resolved.
The silence function works on an alarm-by-alarm basis. An audio alarm will sound
if a new alarm condition occurs after the previous condition has been resolved.
FHR (bpm)
alarm activated
Time (minutes)
100% signal loss
1.25 min
FHR (bpm)
alarm activates
Time (minutes)
70% signal loss
end of “good”
quality signal
5 min

Table of Contents

Other manuals for GE Corometrics 250cx Series

Related product manuals