Revision C 250cx Series Maternal/Fetal Monitor 3-11
2036946-001
Controls, Indicators, and Connectors: MHR/P Area
MHR/P Area
The MHR/P area is summarized by the following figure and table.
MHR/P Display
MHR/P Display
Name Description
A
MHR/P Value Up to three-digits indicate the MHR/P in beats per
minute.
B
MHR/P Alarm Setting
Indicator
This symbol provides information about the MHR/P
audio alarm and the MHR/P high/low alarm limit
settings. See Chapter 10, “Alarms” for more
information. Maternal alarms cannot be disabled.
C
MHR/P Mode Title The mode title MECG indicates MECG is the
MHR/P source; the mode title Pulse indicates
MSpO
2
or NIBP is used as the source. Select the
mode title softkey to access the MHR/P Setup
screen.
D
Maternal Heartbeat
Indicator
Flashes with each detected valid heartbeat—for
MECG only.
C
D
B
A