Revision C 250cx Series Maternal/Fetal Monitor 8-7
2036946-001
Maternal Non-Invasive Blood Pressure Monitoring: NIBP Setup Screen
NIBP Setup Screen
Select the NIBP softkey to access the NIBP Setup screen.
Maternal NIBP Setup Screen
Mode
This field alternates between the manual and automatic monitoring modes for
maternal blood pressure. For auto mode, this field also sets the interval time, in
minutes, between automatic blood pressure determinations. This interval time is
measured from beginning to beginning of determinations. (The monitor is factory-
set with the optional 1-minute interval time enabled. For information on disabling
the 1-minute interval, refer to the “250/250cx Series Monitor Service Manual”.)
NOTE: As soon as the auto mode is selected on the setup screen, the countdown
timer begins to decrement. The first automatic determination begins after
expiration of one complete interval time period.
Target
This option lets you choose the initial pressure for cuff inflation. If the NIBP is taken
while previous determination is still displayed, and within 16 minutes of the
previous determination, the initial target pressures for subsequent determinations are
based upon the systolic values of previous determination. The default initial target
pressure is 135 mmHg (18.0 kPa). Adjust the pressure between 100 to 250 mmHg
(13.3 to 33.3 kPa) in increments of 5 mmHg (0.7 kPa).
NOTE: Selecting a target pressure will clear old NIBP values in the vital signs area
and cancel any determination in progress.
165
172
30
130/ 85
NIBP
MAP (107)
03:15
US
FECG
TOCO
11:41
NIBP Setup
Alarms High Low
Systolic: 160 90 mmHg
Diastolic: 90 50 mmHg
MAP: 140 50 mmHg
MHR/P: 120 50 bpm
Alarm Volume: 5
NIBP Done Vol:
70-240
50-150
70-130
30-120
30-120
35-120
70-150
100-250
Manual
Auto 1 min
Auto 2 min
Auto 3 min
Auto 4 min
Auto 5 min
Auto 10 min
Auto 15 min
Auto 20 min
Auto 30 min
Auto 40 min
Auto 45 min
Auto 60 min
Auto 90 min
Auto 120 min
1
2
3
4
5
6
7
8
9
0
1
2
3
4
5
6
7
8
9
High Low
Target:
100-250,
5 mmHg
increments
Mode:
135
Exit
5
Manual
9.3-32.0
6.7-20.0
9.3-17.3
4.0-16.0
4.0-16.0
35-120
9.3-20.0
100-250
High Low
mmHg mode
kPa mode