Revision C 250cx Series Maternal/Fetal Monitor 7-3
2036946-001
Maternal Heart/Pulse Rate Monitoring: MHR/P Source
MHR/P Source
The MHR/P can be determined by the MECG, MSpO
2
, and NIBP sections of the
monitor. However, the data from only one parameter is:
referred to as the MHR/P source
displayed in the MHR/P area;
used to evaluate an MHR/P alarm condition; and
used to generate the MHR/P trace on the strip chart paper.
1
The source is:
selected via the MHR/P Setup screen
may be manually selected or automatically selected by the monitor according to
the following priority order:
MECG (values updated continuously)
MSpO
2
(values updated continuously)
NIBP (NIBP is available only as an Auto selection; the Manual selection is
disabled. Values updated only when NIBP determinations are taken)
IMPORTANT:
MSpO
2
AS AN MHR/P SOURCE—If MSpO
2
is selected as the
MHR/P source, an MHR/P alarm only occurs if the pulse rate
value derived from the MSpO
2
sensor violates an MHR/P alarm
limit. The MHR/P values derived from the MECG and NIBP
sections of the monitor are ignored. The heart rate tone varies in
pitch to reflect changes in the maternal oxygen saturation reading.
The pitch rises as the saturation values increase, and lowers as the
saturation values decrease. The pulse rate trend is a grey line
annotated by MSpO
2
P .
1
If NIBP is selected as the MHR/P source, there is no trending of the data since these are static
measurements.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals