10-4 250cx Series Maternal/Fetal Monitor Revision C
2036946-001
Alarms: Alarm Setup
The alarm limits for each modality are configured by a respective setup screen.
Refer to Chapter 4, “Setup Procedures”. A Master Alarm Setup screen provides a
summary of most alarm limit settings with the exception of the FHR1 and FHR2
limit settings which are set independently.
NOTE: For each modality, the available ranges of high and low alarm limits
overlap; however, the monitor prevents the selection of overlapping alarm
limits.
Alarm Volume
The alarm volume can be set on any individual setup screen or on the Master Alarm
Setup screen. This settings is used for all alarms.
CAUTION
ALARM SETUP—Prior to monitoring each patient, it is
recommended that you check the alarm limits to ensure they are
appropriate for the patient.
Alarm Silence
The figure below is a sample of the ALARM SILENCE message on-screen. Press the
Alarm Silence button to silence an individual maternal audio alarm or paper load
error. However, for MECG and MSpO
2
monitoring and during a paper-load error
condition, an alarm will be reissued if the alarm state continues after a specified
amount of time. Once alarm silence is activated, an ALARM SILENCE message box
appears on-screen with a timer that counts down the remaining time to re-alarm.