Revision C 250cx Series Maternal/Fetal Monitor 17-3
2036946-001
Technical Specifications: General Monitor
General Monitor
General Monitor Technical Specifications
Category Technical Specifications
Power Requirements
Nominal Line Voltage:
Line Frequency:
Power Consumption (maximum):
Chassis Leakage:
100VAC 120 VAC 220 VAC 230 VAC 240 VAC
50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz
100 W 100 W 0.4 A 0.4 A 0.4 A
<300 µA
Physical Characteristics
Height:
Width:
Depth:
Weight:
6.7 in (17.0 cm)
16.7 in (42.4 cm)
17.5 in (44.4 cm)
22.0 lbs (10.9 kg) approx.
Environmental Conditions
Monitor:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
Strip Chart Paper
1
:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
Operating Storage
50°F to 104°F (10°C to 40°C) 14°F to 131°F (–10°C to 55°C)
10% to 95%, non-condensing 0% to 95%, non-condensing
700–1060 mbar (525–795 mmHg) 700–1060 mbar (525–795 mmHg)
50°F to 104°F (10°C to 40°C) < 80°F (< 26.5°C)
30% to 70%, non-condensing 45% to 65%, non-condensing
700–1060 mbar (525–795 mmHg) 700–1060 mbar (525–795 mmHg)
Certification
ANSI/AAMI EC13-1992:
UL-2601.1:
CUL:
Complies with all areas except those listed below:
3.1.2.1e: Heart Rate Meter Accuracy and Response to Irregular Rhythm (not tested)
3.2.6.1: Range of QRS wave amplitude and duration
3.2.7: Range and accuracy of heart rate meter (4.2.7 f: input rate of 300 bpm.)
3.2.8.1: Lower Alarm Limit (The lowest alarm limit on the 250cx Series is 35 bpm.)
3.2.9.7a: Output Display a) Channel Width
3.2.9.8c: Impulse Response
3.2.9.12: Pacemaker Pulse Display capability
Classified to UL-2601.1
Medical electrical equipment classified by Underwriter’s Laboratories, Inc., with respect to
fire, shock, and mechanical hazards in accordance with UL-2601.1.
Classified with respect to electric shock, fire, mechanical, and other specified hazards
only, in accordance with CAN/CSA C22.2 No. 601.1
1
Paper operating environmental conditions are for a period of less than one month. Paper storage environmental conditions are for extended storage.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD