Revision C 250cx Series Maternal/Fetal Monitor 17-3
2036946-001
Technical Specifications: General Monitor
General Monitor
General Monitor Technical Specifications
Category Technical Specifications
Power Requirements
Nominal Line Voltage:
Line Frequency:
Power Consumption (maximum):
Chassis Leakage:
100VAC 120 VAC 220 VAC 230 VAC 240 VAC
50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz 50/60 Hz
100 W 100 W 0.4 A 0.4 A 0.4 A
<300 µA
Physical Characteristics
Height:
Width:
Depth:
Weight:
6.7 in (17.0 cm)
16.7 in (42.4 cm)
17.5 in (44.4 cm)
22.0 lbs (10.9 kg) approx.
Environmental Conditions
Monitor:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
Strip Chart Paper
1
:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
Operating Storage
50°F to 104°F (10°C to 40°C) 14°F to 131°F (–10°C to 55°C)
10% to 95%, non-condensing 0% to 95%, non-condensing
700–1060 mbar (525–795 mmHg) 700–1060 mbar (525–795 mmHg)
50°F to 104°F (10°C to 40°C) < 80°F (< 26.5°C)
30% to 70%, non-condensing 45% to 65%, non-condensing
700–1060 mbar (525–795 mmHg) 700–1060 mbar (525–795 mmHg)
Certification
ANSI/AAMI EC13-1992:
UL-2601.1:
CUL:
Complies with all areas except those listed below:
3.1.2.1e: Heart Rate Meter Accuracy and Response to Irregular Rhythm (not tested)
3.2.6.1: Range of QRS wave amplitude and duration
3.2.7: Range and accuracy of heart rate meter (4.2.7 f: input rate of 300 bpm.)
3.2.8.1: Lower Alarm Limit (The lowest alarm limit on the 250cx Series is 35 bpm.)
3.2.9.7a: Output Display a) Channel Width
3.2.9.8c: Impulse Response
3.2.9.12: Pacemaker Pulse Display capability
Classified to UL-2601.1
Medical electrical equipment classified by Underwriter’s Laboratories, Inc., with respect to
fire, shock, and mechanical hazards in accordance with UL-2601.1.
Classified with respect to electric shock, fire, mechanical, and other specified hazards
only, in accordance with CAN/CSA C22.2 No. 601.1
1
Paper operating environmental conditions are for a period of less than one month. Paper storage environmental conditions are for extended storage.