14-6 250cx Series Maternal/Fetal Monitor Revision C
2036946-001
Waveforms: Printing a Waveform Snapshot
Recorder in Maternal-Only Mode
If the recorder is running in maternal-only mode, the waveform does not interrupt
the printing of data and prints at a high-speed mode of 12 cm/min. When finished,
the recorder returns to the maternal-only mode.
Recorder Off
If the recorder is off, the waveform print using a high-speed mode of 12 cm/min
after which the recorder turns off again.
NOTE: You cannot switch between different types of waveforms without
cancelling any printing that is in progress.
NOTE: The monitor must collect 6 seconds of new data following completion of a
print function before it can print again.
Stopping a Print Command
The following actions interrupt waveform printing:
changing the recorder mode
pressing the Test button
opening the recorder door
running out of paper
turning off the monitor
CAUTION
DATA STORAGE—Stored data for the maternal Vital Signs History
screen is immediately lost when the monitor is turned off. This
ensures that stored data for one patient is not inadvertently
transferred to a new patient. Refer to Chapter 12, “Maternal Vital
Signs History” for more information.