Revision C 250cx Series Maternal/Fetal Monitor 11-15
2036946-001
Recorder Modes: Chart Style Vital Signs Printing
Examples of Printing Styles
Chart-Style Printing Examples
When chart-style printing is enabled:
The 15-minute interval prints on the quarter hour (e.g.,
9:00, 9:15, 9:30, 9:45,
etc.).
The 30-minute interval prints on the half hour (e.g., 9:00, 9:30, 10:00, 10:30, etc.).
The 60-minute interval prints on the hour (e.g., 9:00, 10:00, 11:00, 12:00, etc.).
The following are examples of chart-style printing:
NIBP Example 1: The automatic blood pressure mode is activated at 9:03, with
the interval time set to 15 minutes. Whereas the first real-time reading would
occur at 9:18, the first chart-style reading is taken at 9:15. Subsequent readings
are taken and printed at 9:30, 9:45, 10:00, 10:15, etc.
NIBP Example 2: The automatic blood pressure mode is activated at 9:17, with
the interval time set to 30 minutes. Whereas the first real-time reading would
occur at 9:47, the first chart-style reading is taken at 9:30. Subsequent readings
are taken and printed at 10:00, 10:30, 11:00, 11:30, etc.
MSpO2 Example 1: The MSpO2 print interval time is set to 30 minutes and the
sensor is connected at 9:24 a.m. Whereas the first real-time printing would
occur at 9:54, the first chart-style printing is done at 9:30. Subsequent values
are printed at 10:00, 10:30, 11:00, 11:30, etc.
MSpO2 Example 2: The MSpO2 print interval time is set to 60 minutes and the
sensor is connected at 9:42 a.m. Whereas the first real-time printing would
occur at 10:42, the first chart-style printing is done at 10:00. Subsequent values
are printed at 11:00, 12:00, 1:00, 2:00, etc.
Real-Time Printing Example
When chart-style printing is disabled, standard real-time printing occurs.
NIBP Example: The automatic blood pressure mode is activated at 9:03, with
the interval time set to 15 minutes. The first reading occurs at 9:18. Subsequent
readings are taken at 9:33, 9:48, 10:03, etc.
MSpO2 Example: The print interval is set to 15 minutes and the first acceptable
pulse signal is detected at 9:05 a.m. The first printing occurs at 9:05.
Subsequent MSpO
2 values are printed at 9:20, 9:35. 9:50, 10:05, etc.
Chart-Style 7-Minute Exception for NIBP
If you take a manual blood pressure reading within 7 minutes of a chart-style
interval (15, 30, or 60 minutes) and then activate automatic blood pressure readings
using a chart-style interval, the first automatic reading will be skipped.
This rule only applies to the first reading when chart-style vital signs printing is
enabled on the password-protected Install Options service screen.
Example: You take a manual blood pressure reading at 7:10. At 7:13 you then
activate automatic blood pressure readings using a 15-minute time interval. The
7:15 chart-style reading is skipped since a manual reading occurred 5-minutes
earlier. The first automatic reading is taken at 7:30. Since the rule applies only
to the first reading, if you take another manual reading at 7:40, the 7:45
automatic reading occurs as scheduled.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals