Revision C 250cx Series Maternal/Fetal Monitor 10-3
2036946-001
Alarms: Introduction
Introduction
This chapter provides a summary of alarms for all modalities in the 250cx Series
monitor. The monitor provides patient alarms (alarm limit threshold violations) for:
FHR1
FHR2
NIBP (systolic, diastolic, and mean arterial pressures)
MHR/P (for the selected source)
MSpO
2
In addition, the monitor provides signal quality alarms.
NOTE: The audio portion of an alarm takes priority to override the song player if
activated.
Alarm Setup
Master Alarm Setup Screen
The figure below is a sample Master Alarm Setup screen. Although each of the fields
on this screen can be accessed under the individual parameter setup screens, the
Master Alarm Setup screen provides an overall summary of the maternal alarm setup
information
.
Master Maternal Alarm Setup Screen
Alarms
These fields adjust the high and low alarm limits for NIBP, MHR/P, and MSpO
2
.
The available ranges are shown the “Technical Specifications” section. The factory
default setting are listed in Appendix A, “Factory Defaults”.
165
172
30
US
FECG
TOCO
Master Alarm Setup
Alarms High Low
Systolic:
160 90 mmHg
Diastolic:
90 50 mmHg
MAP: 140 50 mmHg
MHR/P:
120 50 bpm
MSpO2:
100 95 %
Alarm Volume: 5
Exit