Revision C 250cx Series Maternal/Fetal Monitor C-17
2036946-001
Spectra Alerts: Alert Parameters
Alert Parameters
Summary of Alert Parameters
ALERT PARAMETERS SUMMARY ON FHR SETUP SCREEN TREND SCREEN
Parameter Column 1 Column 2 Message Criteria
Alert Level
*
**
***
*
**
***
*
**
- message
***
Alert has not been silenced.
Alert has been silenced.
Baseline HR
(bpm)
Average rate over
past 10 minutes.
High
BASELINE?
Alert pending: FHR > 160 bpm for 5
minutes.
Alert has not been silenced. Reflects the
detection of baseline FHR > 160 bpm for 10
minutes.
Alert silenced.
Low
BASELINE?
Alert pending. FHR < 120 bpm for 5
minutes.
Alert has not been silenced. Reflects the
detection of baseline FHR < 120 bpm for a
pre-determined period of time. The alert
occurs in 2–10 minutes, depending on how
low the rate goes.
Alert silenced.
Alert is deleted when baseline FHR is within
the “normal” range for 10 minutes.
+ message?
Tachy
Tachy
BASELINE?
Brady
Brady
BASELINE?