5-4 250cx Series Maternal/Fetal Monitor Revision C
2036946-001
Fetal Heart Rate Monitoring: FECG (Internal Method)
Alert
This field controls and shows Spectra Alerts. Refer to Appendix C for more
information.
Alarm Volume
This field controls the alarm volume for all fetal alarms.
FECG (Internal Method)
Methodology
This method uses an electrode attached directly to the fetal presenting part. The
electrode is connected to the cable/legplate secured to the mother. The fetal heart
rate is computed based upon the interval between successive R-wave peaks of the
fetal QRS complex.
The fetal heart rate is displayed in bpm and is continuously plotted on the strip chart
recorder paper if the recorder is on. (Refer to “US/US2 Setup Screen” on page 5-3.)
The heartbeat indicator flashes for each detected heartbeat.
Artifact Elimination
An FECG artifact elimination option is available behind the password-protected
Service Lock screen on all 250cx Series Monitors.

Table of Contents

Other manuals for GE Corometrics 250cx Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Corometrics 250cx Series and is the answer not in the manual?

GE Corometrics 250cx Series Specifications

General IconGeneral
TypeFetal Monitor
Model250cx Series
ManufacturerGE Healthcare
Fetal Heart Rate MonitoringYes
Ultrasound Frequency1 MHz
Recorder Speed1, 2, 3 cm/min
Maternal ParametersECG, NIBP, SpO2
FHR Range30-240 bpm
TOCO Range0-100
ConnectivityRS-232
Power SourceAC Power
DisplayColor LCD

Related product manuals